Humidity Chamber Water: Scale in the Boiler and the 93 % RH That Drifts

2026-11-19

Humidity is produced by boiling water or by atomising it. Both methods have the same long-term problem: whatever is dissolved in the water stays in the machine. Over a year of daily use, a chamber can accumulate enough scale to change how much vapour it can produce at a given heater setting.

What the water carries

Tap water in a typical city carries 150 to 400 mg/L of dissolved solids, mostly calcium and magnesium carbonate. A chamber that evaporates 5 L a day for 200 working days deposits between 150 and 400 g of solids in a year. Some leaves with the drain; the rest stays on the heater and in the boiler.

Water sourceDissolved solidsEffect on the chamber
Tap water150 to 400 mg/LScale in 6 to 12 months
Softened water20 to 80 mg/LSlower scale, sodium salts instead
Deionised waterUnder 5 mg/LMinimal scale, but low conductivity
RO water10 to 30 mg/LGood compromise for most labs

Why scale changes the reading

Scale on a heater acts as an insulator. The heater runs hotter to deliver the same wattage to the water, and the controller, which reads a temperature sensor, compensates by adjusting the duty cycle. At high humidity setpoints the chamber runs out of headroom before the setpoint is reached, and the display slowly stops reaching 93 %.

The second effect is on the sensor itself. A wet-bulb wick that has absorbed hard water dries to a crust and reads low, which makes the chamber over-humidify in an attempt to correct a false reading.

Maintenance that prevents it

  • Use RO water, or at minimum softened water, and record the source on the log.
  • Descale the boiler every 6 months, or every 3 months with hard water.
  • Replace the wet-bulb wick monthly and use deionised water for the reservoir.
  • Check the humidity reading against an independent hygrometer each quarter.

Standards referenced: IEC 60068-2-78 for damp heat steady state, IEC 60068-3-6 for humidity chamber confirmation.

The conductivity question

Deionised water solves the scale problem and creates another: at very low conductivity, some level sensors stop working, and the chamber reports a water fault that is not a fault.

The usual compromise is RO water at 10 to 30 mg/L, or a blend of RO and tap water that keeps conductivity in the 5 to 50 µS/cm range that most sensors handle comfortably.

What the log should contain

Water source, conductivity, date of last descale, date of last wick change, and any humidity deviation observed. Five fields, filled in weekly, turn a slow drift into a visible trend.

The deviation field is the important one. A chamber that reads 1 % low at 93 % RH is within many tolerances; a chamber that reads 1 % low this month and 3 % low in three months is on its way to a fault.

Descaling without damage

Use a citric acid solution at 5 to 10 %, circulate for 30 minutes, rinse twice with RO water, then run a humidification cycle and drain. Do not use hydrochloric acid based descalers on stainless boilers: they pit the surface and create new nucleation sites for scale.

Record the descale in the maintenance log and check the humidity reading against an independent instrument afterwards. A descale that changes the reading by more than 2 % is a sign the scale was significant.