Humidity is produced by boiling water or by atomising it. Both methods have the same long-term problem: whatever is dissolved in the water stays in the machine. Over a year of daily use, a chamber can accumulate enough scale to change how much vapour it can produce at a given heater setting.
Tap water in a typical city carries 150 to 400 mg/L of dissolved solids, mostly calcium and magnesium carbonate. A chamber that evaporates 5 L a day for 200 working days deposits between 150 and 400 g of solids in a year. Some leaves with the drain; the rest stays on the heater and in the boiler.
| Water source | Dissolved solids | Effect on the chamber |
| Tap water | 150 to 400 mg/L | Scale in 6 to 12 months |
| Softened water | 20 to 80 mg/L | Slower scale, sodium salts instead |
| Deionised water | Under 5 mg/L | Minimal scale, but low conductivity |
| RO water | 10 to 30 mg/L | Good compromise for most labs |
Scale on a heater acts as an insulator. The heater runs hotter to deliver the same wattage to the water, and the controller, which reads a temperature sensor, compensates by adjusting the duty cycle. At high humidity setpoints the chamber runs out of headroom before the setpoint is reached, and the display slowly stops reaching 93 %.
The second effect is on the sensor itself. A wet-bulb wick that has absorbed hard water dries to a crust and reads low, which makes the chamber over-humidify in an attempt to correct a false reading.
Standards referenced: IEC 60068-2-78 for damp heat steady state, IEC 60068-3-6 for humidity chamber confirmation.
Deionised water solves the scale problem and creates another: at very low conductivity, some level sensors stop working, and the chamber reports a water fault that is not a fault.
The usual compromise is RO water at 10 to 30 mg/L, or a blend of RO and tap water that keeps conductivity in the 5 to 50 µS/cm range that most sensors handle comfortably.
Water source, conductivity, date of last descale, date of last wick change, and any humidity deviation observed. Five fields, filled in weekly, turn a slow drift into a visible trend.
The deviation field is the important one. A chamber that reads 1 % low at 93 % RH is within many tolerances; a chamber that reads 1 % low this month and 3 % low in three months is on its way to a fault.
Use a citric acid solution at 5 to 10 %, circulate for 30 minutes, rinse twice with RO water, then run a humidification cycle and drain. Do not use hydrochloric acid based descalers on stainless boilers: they pit the surface and create new nucleation sites for scale.
Record the descale in the maintenance log and check the humidity reading against an independent instrument afterwards. A descale that changes the reading by more than 2 % is a sign the scale was significant.