Two chambers both read 93 % RH. One runs at 40 °C, the other at 25 °C. The first has a dew point near 39 °C, the second near 24 °C, and a sample that arrives 5 K colder than the chamber will condense water in one and not the other.
That difference is the whole point of the test. Condensation on an insulating surface creates a leakage path, and leakage is what the damp heat test is designed to provoke.
Dew point is the temperature at which the air is saturated. At 93 % RH it sits about 1 K below the chamber temperature, which is why the sample only has to be 2 K colder than the air for water to form.
| Chamber condition | Dew point | Sample at 30 °C | Sample at 20 °C |
| 40 °C / 93 % RH | About 39 °C | Condenses | Condenses heavily |
| 40 °C / 50 % RH | About 26 °C | No condensation | Condenses |
| 25 °C / 93 % RH | About 24 °C | No condensation | Condenses |
| 25 °C / 50 % RH | About 13 °C | No condensation | No condensation |
This is why a cold sample carried straight from an unheated store into a warm humid chamber can condense water on arrival, before the test has even started. The preconditioning step exists to prevent exactly that.
Record chamber temperature, chamber humidity, dew point and the sample surface temperature. The first two are what the standard asks for; the last two are what tells you whether the test did what it claims. A single thermocouple taped to a sample is enough, and it turns a report from a claim into evidence.
Our constant climate chambers log dew point directly rather than deriving it, and the technical notes cover the recovery times for each chamber size.
Standards referenced: IEC 60068-2-78 (damp heat, steady state), IEC 60068-2-30 (damp heat, cyclic), IEC 60068-3-6.
A 24 h damp heat soak at 40 °C and 93 % RH puts a dew point of about 39 °C against a sample that starts at 25 °C, so condensation is guaranteed in the first hour. That is the intent: the test is looking for the leakage path that only exists when water is present.
Insulation resistance measured at 500 V after the soak, with the sample still wet, is the value that matters; a reading taken after 2 h of drying tells you about a different product.