Technical Articles

Wet-Bulb and Dry-Bulb: The Two Thermometers Behind Every Humidity Reading

2026-11-20

Relative humidity is not measured. It is calculated, from two temperatures, and the calculation is only as good as the wick that connects them. This is why a chamber can pass a calibration in January and read 5 % low in July with the same sensor and the same controller, and why the first thing to check on a drifting humidity reading is not the electronics.

The two temperatures

The dry-bulb thermometer measures the air temperature. The wet-bulb thermometer measures the lowest temperature reachable by evaporation from a wetted wick in that air. The difference between them, the wet-bulb depression, carries the humidity information: a large depression means dry air, a small one means humid air, and zero means the air is saturated.

Everything depends on the evaporation being unimpeded. A wick that has scaled up, a wick that has gone hard, a reservoir that has run dry, or a fan that has stopped all reduce evaporation, and each of them pushes the wet-bulb reading toward the dry-bulb reading, which the controller interprets as higher humidity. A chamber can sit at a displayed 95 % RH while the working volume is at 88 %.

FaultWet-bulb readingDisplayed RHActual RH
Healthy wick, distilled waterCorrectCorrectCorrect
Scaled or hardened wickToo highToo high by 3 to 8 %Lower than displayed
Reservoir emptyEquals dry bulbNear 100 %Unknown, likely far lower
Fan stoppedToo highToo high by 5 to 15 %Much lower
Tap water with 300 µS/cmDrifts over weeksSlowly too highLower than displayed
Wick installed but not wettedEquals dry bulbNear 100 %Uncontrolled

Water quality is not a detail

The wick is a distillation column running backwards. Water evaporates from it and leaves behind everything that was dissolved in it, so hardness, silica and chloride concentrate in the fabric until the pores close. With tap water at 300 µS/cm, a wick can go from serviceable to useless in three to six weeks of continuous running. With distilled or deionised water below 10 µS/cm, the same wick lasts three to six months.

The reservoir is the other half of the problem. A tank that is topped up but never drained accumulates salts indefinitely; the concentration only goes up. A monthly drain and refill costs ten minutes and removes the mechanism entirely.

Checks that take under five minutes

  • Wick colour and stiffness: a wick that has gone brown or feels crisp when pinched is finished, whatever its age.
  • Wick contact: it must wrap the bulb fully and sit in the water, not above it.
  • Reservoir level and water conductivity, if a meter is available; above 10 µS/cm is a finding.
  • Fan operation, felt as a steady draught at the wet bulb, not guessed from the noise.
  • Agreement between the chamber display and an independent logger at the working volume, at one humidity point, monthly.

Why the independent logger earns its price

A chamber that calculates humidity from its own wet bulb has no way to notice that the wet bulb is wrong: the fault is inside the measurement chain, and the controller trusts the chain. An independent capacitive logger at the working volume is a second, differently-failing instrument, and that is the whole point. When the two agree at 40 °C and 93 % RH within 2 %, both are probably right; when they diverge by 5 %, one of them has a reason, and the wet bulb is the usual suspect.

For steady-state damp heat work to IEC 60068-2-78 the tolerance on relative humidity is tight enough that a 5 % error changes the conclusion of a 1,000 hour test. That is a large consequence for a consumable item that costs less than a coffee, which is roughly the argument to put in the maintenance budget request.

Standards referenced: IEC 60068-2-78 (damp heat, steady state), IEC 60068-2-30 (damp heat, cyclic), IEC 60068-3-6 (confirmation of the performance of temperature/humidity chambers).

close
Scan the qr codeClose
the qr code