Probe 18 and Probe 19: Two Child Probes, Two Age Groups, One Mistake

2026-11-26

Two probes in the IEC 61032 range represent a child's finger. They look similar, they are both jointed, and they are frequently confused. They model different ages, and the difference decides which openings a product must resist.

The dimensions

ProbeFingerLengthAge
188.6 mm, R4.357.9 mm, three segments36 months to 14 years
195.6 mm, R2.844 mm, three segmentsup to 36 months

Probe 18 is the larger of the two. Its 8.6 mm diameter and 57.9 mm length come from measurements of children from about three years upward. Probe 19 is smaller in both dimensions because a toddler's finger is shorter as well as thinner.

Why both exist

A product intended for a three-year-old and a product intended for a ten-year-old present different hazards. The older child can reach further into an opening and can apply more force; the younger one reaches less far but will push a finger into anything.

Using probe 18 on a product for toddlers produces a test that under-estimates reach: the probe is too long and too thick to enter openings a toddler's finger would enter.

Using probe 19 on a product for older children produces the opposite: it enters openings the older child's finger cannot, and generates failures that do not exist.

The handle is part of the definition

On probe 19 the handle is not simply a grip. It has two sections of defined diameter and length, and those dimensions affect how deep the finger can be pressed. The transition from finger to handle acts as a stop, and a probe assembled with the wrong handle tests to the wrong depth.

This is a common failure on probes that have been repaired. A handle replaced with a generic one changes the effective depth without changing any of the finger dimensions that are usually checked.

What to check at inspection

For both probes:

Finger diameter at each of the three segments. Wear is not uniform; the tip segment wears first.

Total length and segment lengths. A joint that has been forced can change the segment lengths without changing the overall appearance.

Joint freedom and play. A stiff joint will not follow a contour; a loose one will sag and reach further than it should.

For probe 19, the two handle sections. Diameter and length, because they define the stop.

Recording the choice

The test report should name the probe and the age group it represents. "Child finger probe" alone is ambiguous, and the ambiguity matters when a customer asks why a product passed one standard and failed another.

The age group is the useful field. It tells the reader which population the finding applies to, which is what the standard is actually about.